beautypg.com

Technical data – KROHNE OPTIBAR DP 7060 C EN User Manual

Page 66

background image

7

TECHNICAL DATA

66

OPTIBAR DP 7060 C

www.krohne.com

04/2014 - 4003292001 - MA OPTIBAR DP 7060 C R01 en

Effect of ambient

temperature

Ambient temperature effect on zero and span.

[% of URL]

up to nominal

pressure acc. to

URL absolute

pressure

-10...+60°C /

+14...140°F

-40...+80°C / -40...+176°F

30 mbar / 0.44 psi

40 bar / 580 psi

<± 0.10%

<± 0.15%

100 mbar / 1.5 psi

160 bar / 2320 psi

500 mbar / 7.3 psi

3 bar / 43.51 psi

16 bar / 232.1 psi

Long-term stability

acc. to

DIN EN 61298-1

<± 0.1% of URL over a period of 5 years

Operating conditions

Temperature

Temperature

Temperature

Temperature

Operating

temperature / nominal

temperature range

-40...+80°C / -40...+176°F

Ambient temperature

-40...+80°C / -40...+176°F

Storage temperature

-40...+80°C / -40...+176°F

Climate category

4K 4H (air temperature: -20...+55°C, humidity: 4...100% according to DIN EN 60721-3-4)

Further operating conditions

Ingress protection acc. to IEC 529 / EN 60529

Ingress protection acc. to IEC 529 / EN 60529

Ingress protection acc. to IEC 529 / EN 60529

Ingress protection acc. to IEC 529 / EN 60529

Plastic (PBT)

Single chamber

IP66 / IP67

NEMA 4X / 6P

Double chamber

IP66 / IP67

NEMA 4X

Aluminium

Single chamber

IP66 / IP68 (0.2 bar / 2.9 psi)

NEMA 4X / 6P

IP 68 (1 bar / 14.5 psi)

-

Double chamber

IP66 / IP67

NEMA 4X

IP66 / IP68 (0.2 bar / 2.9 psi)

NEMA 4X

Stainless steel

(electro-polished)

Single chamber

IP69K

-

Single chamber

IP66 / IP68 (0.2 bar / 2.9 psi)

NEMA 4X / 6P

Stainless steel

(precision casting)

Single chamber

IP66 / IP68 (0.2 bar / 2.9 psi)

NEMA 4X / 6P

IP 68 (1 bar / 14.5 psi)

NEMA 4X

Double chamber

IP66 / IP67

NEMA 4X

IP66 / IP68 (0.2 bar / 2.9 psi)

NEMA 4X

Stainless steel

Sensor for external housing

IP 68 (25 bar / 362.6 psi)

-

.book Page 66 Friday, May 2, 2014 10:25 AM