TECHNICAL DATA
66
OPTIBAR DP 7060 C
www.krohne.com
04/2014 - 4003292001 - MA OPTIBAR DP 7060 C R01 en
Effect of ambient
temperature
Ambient temperature effect on zero and span.
[% of URL]
up to nominal
pressure acc. to
URL absolute
pressure
-10...+60°C /
+14...140°F
-40...+80°C / -40...+176°F
30 mbar / 0.44 psi
40 bar / 580 psi
<± 0.10%
<± 0.15%
100 mbar / 1.5 psi
160 bar / 2320 psi
500 mbar / 7.3 psi
3 bar / 43.51 psi
16 bar / 232.1 psi
Long-term stability
acc. to
DIN EN 61298-1
<± 0.1% of URL over a period of 5 years
Operating conditions
Temperature
Temperature
Temperature
Temperature
Operating
temperature / nominal
temperature range
-40...+80°C / -40...+176°F
Ambient temperature
-40...+80°C / -40...+176°F
Storage temperature
-40...+80°C / -40...+176°F
Climate category
4K 4H (air temperature: -20...+55°C, humidity: 4...100% according to DIN EN 60721-3-4)
Further operating conditions
Ingress protection acc. to IEC 529 / EN 60529
Ingress protection acc. to IEC 529 / EN 60529
Ingress protection acc. to IEC 529 / EN 60529
Ingress protection acc. to IEC 529 / EN 60529
Plastic (PBT)
Single chamber
IP66 / IP67
NEMA 4X / 6P
Double chamber
IP66 / IP67
NEMA 4X
Aluminium
Single chamber
IP66 / IP68 (0.2 bar / 2.9 psi)
NEMA 4X / 6P
IP 68 (1 bar / 14.5 psi)
-
Double chamber
IP66 / IP67
NEMA 4X
IP66 / IP68 (0.2 bar / 2.9 psi)
NEMA 4X
Stainless steel
(electro-polished)
Single chamber
IP69K
-
Single chamber
IP66 / IP68 (0.2 bar / 2.9 psi)
NEMA 4X / 6P
Stainless steel
(precision casting)
Single chamber
IP66 / IP68 (0.2 bar / 2.9 psi)
NEMA 4X / 6P
IP 68 (1 bar / 14.5 psi)
NEMA 4X
Double chamber
IP66 / IP67
NEMA 4X
IP66 / IP68 (0.2 bar / 2.9 psi)
NEMA 4X
Stainless steel
Sensor for external housing
IP 68 (25 bar / 362.6 psi)
-
.book Page 66 Friday, May 2, 2014 10:25 AM