START-UP

5

49

OPTIBAR DP 7060 C

www.krohne.com

04/2014 - 4003292001 - MA OPTIBAR DP 7060 C R01 en

5.4.3 Diagnosis

Device status

Status signals

Check function, Out of Specification, Maintenance required, Failure

Peak value

The respective minimum and maximum pressure values are stored in the device. Under "Peak

values", these values can be viewed or reset. In addition to the pressure, the minimum and maximum

values of the sensor cell and the electronics temperature is stored. These can be viewed or reset

here.

Peak value

"Pressure"

Reset peak value

Peak value

"Differential pressure"

Reset peak value

Peak value

"Static pressure"

Reset peak value

Peak value

"Meas. cell temp."

Reset peak value

Peak value

"Electronics temperature"

Reset peak value

Simulation

In menu item "Simulation", measured values can be simulated via the current output. These are
issued as both analogue and digital (via HART

®

). The simulation is automatically cancelled 60

minutes after the last key stroke.

For applications

"Differential pressure"

Differential pressure, Static pressure, Percent, Current output,

Linear percent, Meas. cell temp. and Electronics temperature

For applications

"Flow"

Flow, Differential pressure, Static pressure, Pressure, Scaled,

Current output, Linear percent, Meas. cell temp. and Electronics

temperature

For applications

"Level"

Filling height or Pressure (Adjustment units), Static pressure,

Percent, Current output, Linear percent, Meas. cell temp. and

Electronics temperature

For applications

"Interface"

Interface height, Differential pressure, Percent, Current output,

Linear percent, Meas. cell temp. and Electronics temperature

For applications

"Density"

Density, Differential pressure, Percent, Current output, Linear

percent, Meas. cell temp. and Electronics temperature