B.99 SYSTem:REPLY? QUERY
SYST:REPLY?
Syntax:
Short Form: SYST:REPLY?
Long Form: SYSTem:REPLY?
Returns:
Description: Indicates whether error message reporting is on (1) or off (0).
Caution: When the GPIB/LAN interface is installed, SYST:REPLY must be on (1). Turning
SYST:REPLY off (0) will cause all commands to take approximately 1 second to complete, and the
responses to queries will have multiple CR LF pairs in front of the actual response.
B.100 SYSTem:SERNumber? QUERY
SYST:SERN?
Syntax:
Short Form: SYST:SERN?
Long Form: SYSTem:SERNumber?
Returns:
Description: Returns unit serial number.
B.101 SYSTem:SETTings? QUERY
SYST:SETT?
Syntax:
Short Form: SYST:SETT?
Long Form: SYSTem:SETTings?
Returns:
Description: Display system settings (see sample response shown below).
B.102 SYSTem:VERSion? QUERY
SYST:VERS?
Syntax:
Short Form: SYST:VERS?
Long Form: SYSTem:VERSion?
Returns:
Description: Identify unit and firmware version.
SYST:SETT?
KEPCO, EL 5K-200-600 01-18-2010,A999104,MCB #200 1.176 $ 2010/01/18 22:27:06 $
Chassis Serial Number: A999104
Voltage range: 400V
Current range: 595.5
System damping: 5
PFM damping: 3
Overvoltage limit: 799.0
Undrvoltage limit: 0.000
Overcurrent limit: 600.0
Overpower limit: 6.000
Max volt limit: 220.0
Max curr limit: 630.0
Max powr limit: 5.250
System mode (startup): CURRent
System mode (now): CURRent
Setpoints
Voltage : 220.0
Current : 0.000
Power : 0.000
Resistance : 1000.
Conductance : 0.255
Fan temperature limits (C)
Lower: 40
Upper: 70
IP Mode : Static
IP Address: 0.0.0.0
Sub Net : 0.0.0.0
Gateway : 0.0.0.0
DNS Addr : 0.0.0.0
syst:vers?
KEPCO, EL 5K-200-600 01-18-1994,A999104,MCB #200 1.176 $ 2010/01/18 22:27:06 $