B.47 [SOURce:]POWer:PROTection COMMAND

POW:PROT

Syntax:

Short Form: [SOUR:]POW:PROT

Long Form: [SOURce:]POWer:PROTection

where = power protection setpoint

Description: Selects power protection setpoint. This is a user-defined value which, if exceeded during operation,

causes a FAULT condition, forcing the load to disengage.

B.48 [SOURce:]POWer:PROTection? QUERY

POW:PROT?

Syntax:

Short Form: [SOUR:]POW:PROT?

Long Form: [SOURce:]POWer:PROTection?

Returns = power protection setpoint

Description: Indicates power protection setpoint.

B.49 [SOURce:]POWer:PROTection:STATe[:LEVel] COMMAND

POW:PROT:STAT

Syntax:

Short Form: [SOUR:]POW:PROT:STAT[:LEV]
Long Form: [SOURce:]POWer:PROTection:STATe[:LEVel]
where: = 1 (enabled) or 0 (disabled)

Description: Clears or sets power protection trip.

B.50 [SOURce:]POWer:PROTection:STATe[:LEVel]? QUERY

POW:PROT:STAT?

Syntax:

Short Form: [SOUR:]POW:PROT:STAT[:LEV]?
Long Form: [SOURce:]POWer:PROTection:STATe[:LEVel]?
where: = 1 (enabled) or 0 (disabled)

Description: Indicates whether the power protection level set using POW:PROT is enabled (1) or disabled

(0).

B.51 [SOURce:]RESistance[:LEVel][:IMMediate][:AMPlitude] COMMAND

RES

Syntax:

Short Form: [SOUR:]RES[:LEV][:IMM][:AMP]
Long Form: [SOURce:]RESistance[:LEVel][:IMMediate][:AMPlitude]
Where: = resistance setpoint

Description: Establishes resistance setpoint. The resistance setpoint only affects the load when the load is in

Resistance mode of operation. Resistance setpoint is stored in non-volatile memory; value is restored
upon power-up.

B.52 [SOURce:]RESistance[:LEVel][:IMMediate][:AMPlitude]? QUERY

RES?

Syntax:

Short Form: [SOUR:]RES[:LEV][:IMM][:AMP]?
Long Form: [SOURce:]RESistance[:LEVel][:IMMediate][:AMPlitude]?
Returns: = resistance setpoint

Description: Indicates resistance setpoint.

B.53 [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude] COMMAND

VOLT

Syntax:

Short Form: [SOUR:]VOLT[:LEV][:IMM][:AMP]
Long Form: [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude]
where: = voltage setpoint

Description: Establishes voltage setpoint. The voltage setpoint only affects the load when the load is in Voltage

mode of operation. Voltage setpoint is stored in non-volatile memory; value is restored upon power-
up.

B.54 [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude]? QUERY

VOLT?

Syntax:

Short Form: [SOUR:]VOLT[:LEV][:IMM][:AMP]?
Long Form: [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude]?
Returns: = voltage setpoint

Description: Indicates voltage setpoint.