beautypg.com

B.32 [source:]measure[:scalar]:current[:dc]? query, B.33 [source:]measure[:scalar]:power[:dc]? query, B.34 [source:]measure[:scalar]:voltage[:dc]? query – KEPCO EL Series Electronic Load Operator Manual P/N 243-1295 Firmware Version 3.22 through 3.41 User Manual

Page 66: B.32, Source, B.33, B.34, Source:]mea, B.35, B.36

background image

B-6

SERIES EL 062011

B.32 [SOURce:]MEASure[:SCALar]:CURRent[:DC]? QUERY

MEAS:CURR?

Syntax:

Short Form: [SOUR:]MEAS[:SCAL]:CURR[:DC]?
Long Form: [SOURce:]MEASure[:SCALar]:CURRent[:DC]?
Returns: Measured values.

Description: Measures current. Measured value returned as XXX Amps, where XXX is decimal data representing

measured current.

B.33 [SOURce:]MEASure[:SCALar]:POWer[:DC]? QUERY

MEAS:POW?

Syntax:

Short Form: [SOUR:]MEAS[:SCAL]:POW[:DC]?
Long Form: [SOURce:]MEASure[:SCALar]:POWer[:DC]?
Returns: Measured values in the form 0.000 Kilowatts

Description: Measures power. Measured value returned as XXX Kilowatts, where XXX represents measured

power.

B.34 [SOURce:]MEASure[:SCALar]:VOLTage[:DC]? QUERY

MEAS:VOLT?

Syntax:

Short Form: [SOUR:]MEAS[:SCAL]:VOLT[:DC]?
Long Form: [SOURce:]MEASure[:SCALar]:VOLTage[:DC]?
Returns: Measured values in the form XXX Volts

Description: Measures voltage. Measured value returned as XXX Volts, where XXX is decimal data representing

measured voltage.

B.35 [SOURce:]CONDuctance[:LEVel][:IMMediate][:AMPlitude] COMMAND

COND

Syntax:

Short Form:[SOUR:]CONDuctance[:LEVel][:IMMediate][:AMPlitude]
Long Form: [SOURce:]CONDuctance[:LEVel][:IMMediate][:AMPlitude]
Where = conductance setpoint in Siemens

Description: Establishes conductance setpoint. Conductance setpoint affects the load only when the load is in

Conductance mode of operation. Conductance setpoint is stored in non-volatile memory; value is
restored upon power-up.

B.36 [SOURce:]CONDuctance[:LEVel][:IMMediate][:AMPlitude]? QUERY

COND?

Syntax:

Short Form:[SOUR:]CONDuctance[:LEVel][:IMMediate][:AMPlitude]?
Long Form: [SOURce:]CONDuctance[:LEVel][:IMMediate][:AMPlitude]?
Returns = conductance setpoint in Siemens

Description: Indicates conductance setpoint.

B.37 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude] COMMAND

CURR

Syntax:

Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP]
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]
Where = current setpoint

Description: Current Setting. The current setpoint affects the load only when the load is in Current mode of opera-

tion. Current setpoint is stored in non-volatile memory; value is restored upon power-up.

B.38 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]? QUERY

CURR?

Syntax:

Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP]?
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]?
Returns = current setpoint

Description: Indicates Current Setpoint.