B.39 [SOURce:]CURRent:PROTection[:LEVel] COMMAND

CURR:PROT

Syntax:

Short Form: [SOUR:]CURR:PROT[:LEV]
Long Form: [SOURce:]CURRent:PROTection[:LEVel]
where: = current protection setpoint

Description: Establishes current protection setpoint. Current protection setpoint is stored in non-volatile mem-

ory; value is restored upon power-up. This is a user-defined value which, if exceeded during opera-
tion, causes a FAULT condition, forcing the load to disengage.

B.40 [SOURce:]CURRent:PROTection[:LEVel]? QUERY

CURR:PROT?

Syntax:

Short Form: [SOUR:]CURR:PROT[:LEV]?
Long Form: [SOURce:]CURRent:PROTection[:LEVel]?
Returns: = current protection setpoint

Description: Indicates current protection setpoint.

B.41 [SOURce:]CURRent:PROTection:STATe COMMAND

CURR:PROT:STAT

Syntax:

Short Form: [SOUR:]CURR:PROT:STAT
Long Form: [SOURce:]CURRent:PROTection:STATe
where = 1 (enabled) or 0 (disabled)

Description: Enables (1) or disables (0) the current protection level set using CURR:PROT. This command

saves having to change the protection level to some value outside of the operating range.

B.42 [SOURce:]CURRent:PROTection:STATe? QUERY

CURR:PROT:STAT?

Syntax:

Short Form: [SOUR:]CURR:PROT:STAT?
Long Form: [SOURce:]CURRent:PROTection:STATe?
Returns: = 1 (enabled) or 0 (disabled)

Description: Indicates whether the current protection level set using CURR:PROT is enabled (1) or disabled

(0).

B.43 [SOURce:]MODE COMMAND

MODE

Syntax:

Short Form: [SOUR:]MODE

Long Form: [SOURce:]MODE

where: = CURRent, POWer, VOLTage, RESistance, CONDuctance, SHORT or OFF.

Description: Sets operating mode: When the operating mode is changed, the input is set to off. Send INPut ON to

apply the new mode setting to the load.

B.44 [SOURce:]MODE? QUERY

MODE?

Syntax:

Short Form: [SOUR:]MODE?

Long Form: [SOURce:]MODE?

Returns: = CURRent, POWer, VOLTage, RESistance, CONDuctance, SHORT or OFF.

Description: Indicates operating mode.

B.45 [SOURce:]POWer[:LEVel][:IMMediate][:AMPlitude] COMMAND

POW

Syntax:

Short Form: [SOUR:]POW[:LEV][:IMM][:AMP]
Long Form: [SOURce:]POWer[:LEVel][:IMMediate][:AMPlitude]
Where = power setpoint

Description: Establishes power setpoint. The power setpoint only affects the load when the load is in Power

mode of operation.

B.46 [SOURce:]POWer[:LEVel][:IMMediate][:AMPlitude]? QUERY

POW?

Syntax:

Short Form: [SOUR:]POW[:LEV][:IMM][:AMP]?
Long Form: [SOURce:]POWer[:LEVel][:IMMediate][:AMPlitude]?
Returns = power setpoint

Description: Indicates power setpoint.