B.23 [SOURce:]INPut[:STATe]? QUERY

INP?

Syntax:

Short Form: [SOUR:]INP[:STAT]?

Long Form: [SOURce:]INPut[:STATe]?

Returns = 0 or OFF (load disengaged) or 1 or ON (load engaged)

Description: Indicates whether load is engaged or disengaged from the UUT. When set to 1 the load is

engaged, when set to 0 the load is disengaged.

B.24 [SOURce:]INPut:PROTection:CLEar COMMAND

INP:PROT:CLE

Syntax:

Short Form: [SOUR:]INP:PROT:CLE

Long Form: [SOURce:]INPut:PROTection:CLEar

Description: Resets protection from overcurrent, overpower, etc.

B.25 [SOURce:]INPut:PROTection:SHORt[:STATe] COMMAND

INP:PROT:SHOR

Syntax:

Short Form: [SOUR:]INP:PROT:SHOR[:STAT]
Long Form: [SOURce:]INPut:PROTection:SHORt[:STATe]

Description: Applies maximum drive to FETs, effectively shorting the input.

B.26 [SOURce:]OUTPut[:STATe] COMMAND

OUTP

Syntax:

Short Form: [SOUR:]OUTP[:STAT] Long Form: [SOURce:]OUTPut[:STATe]
where = 0 or OFF (load disengaged) or 1 or ON (load engaged)

Description: Used to engage or disengage the load from the UUT. When set to 1 the load engages, when set to

0 the load disengages.

B.27 [SOURce:]OUTPut[:STATe]? QUERY

OUTP?

Syntax:

Short Form: [SOUR:]OUTP[:STAT]?

Long Form: [SOURce:]OUTPut[:STATe]?

Returns = 0 or OFF (load disengaged) or 1 or ON (load engaged)

Description: Indicates whether load is engaged or disengaged from the UUT. When set to 1 the load is

engaged, when set to 0 the load is disengaged.

B.28 [SOURce:]OUTPut:PROTection:CLEar COMMAND

OUTP:PROT:CLE

Syntax:

Short Form: [SOUR:]OUTP:PROT:CLE

Long Form: [SOURce:]OUTPut:PROTection:CLEar

Description: Resets protection from overcurrent, overpower, etc.

B.29 [SOURce:]OUTPut:PROTection:SHORt[:STATe] COMMAND

OUTP:PROT:SHOR

Syntax:

Short Form: [SOUR:]INP:PROT:SHOR[:STAT]
Long Form: [SOURce:]INPut:PROTection:SHORt[:STATe]

Description: Applies maximum drive to FETs, effectively shorting the output.

B.30 [SOURce:]MEASure:ALL? QUERY

MEAS:ALL?

Syntax:

Short Form: [SOUR:]MEAS:ALL?

Long Form: [SOURce:]MEASure:ALL?

Returns: Measured values in the form 0.000 KW, 0.000 Amps, 0.000 Volts.

Description: Measures output power (Kilowatts), current (Amps), and voltage (Volts). Unit of measurement

included in the reply.

B.31 [SOURce:]MEASure:ALL2? QUERY

MEAS:ALL2?

Syntax:

Short Form: [SOUR:]MEAS:ALL2?

Long Form: [SOURce:]MEASure:ALL2?

Returns: Measured values in the form 0.000 0.000 0.000 (Kilowatts Amps Volts).

Description: Measures output power (Kilowatts), current (Amps), and voltage (Volts). Unit of measurement is omit-