HT instruments SIRIUS89N User Manual
Page 62

SIRIUS89N
EN - 60
7.2.2. RECORDING: setting of Typical Configurations
The following “
Typical Configurations” are selectable inside meter:
Standard Configuration
Description
LEAKAGE (I1)
Setting of measuring and recording mode of
Leakage current on I1 channel.
TEMP. °C(I1)
Setting of measuring and recording mode of
Temperature (°C) on I1 channel.
HUM %HR(I1)
Setting of measuring and recording mode of
Relativity Humidity %HR on I1
channel.
To activate the above configurations, follow this procedure:
1. Turn the rotary switch on “
AUX”.
2. Press
MENU key. The instrument shows the following screen:
MENU GENERAL
SAFETY TEST MEMORY
ANALYZER MEMORY
RESET
ANALYZER CONFIG
RECORDER CONFIG
CONTRAST
DATE&TIME
LANGUAGE
↓
↑
3. Press
MENU key again. The instrument displays a screen as shown below (picture on
the left). In this way is possible select the desired configuration on I1 channel with
F1 or
F2 key.
TYPICAL CONFIG.
LEAKAGE(I1)
TEMP.°C(I1)
HUM %HR(I1)
↓
↑
TYPICAL CONFIG.
LEAKAGE (I1)
TEMP.°C(I1)
HUM %HR(I1)
Data saved
↓
↑
Configuration selection
Confirmation configuration selection
4. Press
ENTER key. The instrument displays the message “Data saved” for a while to
confirm the desired configuration (see above picture on the right). The instrument
return on the measure mode and you can start recording by pressing
START/STOP
key.