HT instruments SIRIUS89N User Manual
Page 105

SIRIUS89N
EN - 103
14.2. STANDARDS
14.2.1. General
Safety:
EN 61010-1 + A2 (1997)
Protection classification:
Double insulation
Pollution degree:
2
Degree of Protection:
IP50
Over-Voltage Category:
CAT II 600V~ / 350V~ (phase –earth)
CAT III 600V~ / 300V~ (phase –earth)
Usage:
Indoor; max height 2000m
EMC:
EN61326-1 (1997) + A1 (1998) + A2 (2001) + A3 (2003)
The Instrument comply with European Guidelines for CE mark
14.2.2. Safety Test
LOW
Ω (200mA):
IEC 61557-4
M
Ω: IEC
61557-2
RCD: IEC
61557-6
LOOP P-P, P-N, P-PE:
IEC 61557-3
PHASE SEQUENCE:
IEC 61557-7
EARTH: IEC
61557-5
14.2.3. ANALYZER
Voltage sag and surge:
EN50160
Alternating current static Watt-hour meters for active energy:
EN61036 (CLASS 2)
Alternating current static VAR-hour meters for reactive energy: IEC1268 (CLASS 3)
14.2.4. AUX
Sound Level Measurement (with External Probe):
EN 60651:1994/A1 type 1
EN 60804:1994/A2 type 1
14.3. GENERAL
SPECIFICATIONS
14.3.1. Mechanical Data
Dimensions:
225 (L)x165 (W) x 105 (H)mm
Weight: 1,2Kg
approx
14.3.2. Power supply
Batteries:
6 x 1.5-LR6-AA-AM3-MN 1500
Battery Life:
LOW
Ω:
approx: 800 test
M
Ω:
approx: 500 test
RCD AC and A Type:
approx: 1000 test
LOOP P-P, P-N, P-PE
approx: 1000 test
Ra
:
approx: 1000 test
EARTH:
approx: 1000 test
PHASE SEQUENCE:
approx: 1000 test
AUX (recording):
approx: 20 Hours
ANALYZER (recording):
approx: 20 Hours
External power supplier (optional):
Code A0050 or A0053 only for ANALYZER and AUX
functions