HT instruments SIRIUS89N User Manual
Page 14

SIRIUS89N
EN - 12
5.4. RESET
This option re-establishes the initial settings of the instrument in
ANALYZER function.
The “Current Range” parameter it is not modified by Reset command
The initial settings of the instrument consist of:
ANALYZER CONFIG:
System: SINGLE
Frequency: not
modified
Current range:
not modified
Clamp type:
not modified
Transforming ratio of voltmetric transformers:
1
Password: OFF
RECORDER CONFIG:
Start:
MANU (the recording starts at
00 sec mark on clock after
pressing the START/STOP key)
Stop: MAN
Integration period:
15min
Recording of harmonics:
ON
Recording of Voltage anomalies (Sag and Surge:
ON
Voltage Reference for Sag and Surge detection:
230V
Upper Limit for Sag and Surge detection:
6%
Lower Limit for Sag and Surge detection:
10%
Selected voltages:
V1
Selected voltage harmonics:
THD, 01, 03, 05, 07
Selected currents:
I1
Selected current harmonics:
THD, 01, 03, 05, 07
CO-GENERATION: OFF
Powers, Pf and cos
ϕ selected:
P1
Q1i
Q1c
S1
Pf1
DPf1
Energie: Ea1
Eri1
Erc1
The RESET command will not erase the instrument’s memory.