beautypg.com

Measure:frequency, Measure:nwidth, Measure:pduty – GW Instek GDS-1000-U Series Programming Manual User Manual

Page 32

background image

GDS-1000-U Programming Manual

32

:MEASure:FREQuency

Query

Description

Returns the frequency value.
Same as: Measure key → F1~F5 → F3 (Frequency)

Syntax

< Long >

< Short >

:measure:frequency?

:meas:freq?

Returns

Note

Before using this command, select the
measurement channel. See the example below.

Example

:measure:source 1
:measure:frequency?

Selects Channel 1, and
then measures the
frequency.

:MEASure:NWIDth

Query

Description

Returns the first negative pulse width timing.
Same as: Measure key → F1~F5 → F3 (–Width)

Syntax

< Long >

< Short >

:measure:nwidth?

:meas:nwid?

Returns

Note

Before using this command, select the
measurement channel. See the example below.

Example

:measure:source 1
:measure:nwidth?

Selects Channel 1, and
then measures the
negative pulse width.

:MEASure:PDUTy

Query

Description

Returns the positive duty cycle ratio.
Same as: Measure key → F1~F5 → F3 (DutyCycle)

Syntax

< Long >

< Short >