beautypg.com

System:error, System:version – GW Instek GDS-1000-U Series Programming Manual User Manual

Page 13

background image

COMMAND DETAILS

13

*RST

Set

Description

Resets the GDS-1000-U (recalls the default panel
settings).
Same as: Save/Recall key → F1

Syntax

*rst

:SYSTem:ERRor

Query

Description

Returns the oscilloscope system error messages, if
there are any.

Syntax

< Long >

< Short >

:system:error?

:syst:err?

Parameter

ID

Contents

ID

Contents

-100

command error

-102

syntax error

-220

parameter error

-221

settings conflict

-222

data out of range -223 too much data

-224

illegal parameter -232 invalid format

Example

:system:error?
-102

Indicates that the
command syntax is
wrong

:SYSTem:VERSion

Query

Description

Returns the oscilloscope firmware version.
Same as: Utility key → F4 (only the firmware
version)

Syntax

< Long >

< Short >

:system:version?

:syst:vers?

Note

For retrieving all system information including the
firmware version, use the *idn? command.