beautypg.com

Enable period, Delay start point – GW Instek GLA-1000 Series User Manual User Manual

Page 59

background image

Trigger Setup

59

Enable Period

This options controls when GLA actually captures data, in

reference to the Enable Qualifier condition set.

Check the Enable Delay radio button.

According to

GLA captures signals when input signals

match the Enable Qualifier.

Opposite of

GLA captures signals when input signals

DO NOT match the Enable Qualifier.

Delay Start Point This function fine-tunes the length of captured waveform, in

reference to the Enable Qualifier condition. Delay Time configures
the length of time (user-defined). Three types of delay setting,

Start Edge, End Edge, and Period+Delay are available.

Example Enable Qualifier
A0 ~ A3 condition is set as follows.

Enable Qualifier condition: 0001, Delay Time: 100us

Start Edge

GLA starts recording (capturing) signals at

the beginning of Enable Qualifier match,
for the Delay Time duration.

100us

A0

A1

A2

A3

Match

End

Match

Start

Capture

End

Capture

Start