Bus/Signal Setup

41

Clock source

Internal
(asynchronous)
clock

Toolbar icon

Frequency is selectable from 18 frequency

range.

Range: 100Hz ~ 200MHz.

External
(synchronous)
clock

The clock signal needs to be connected to the
CLK input terminal.

Range: 0.001Hz ~ 100MHz.

Rising edge or falling edge is selectable.

Sampling

frequency

To ensure accurate measurement, choose a sampling frequency at
least four times that of the target signal.

Sampling RAM
size

Toolbar icon

Select the amount of waveform memory GLA captures in a single

run. Note that large memory size slows down display update rate.

RAM size

Channel

Compression Enable mode

2k bits

16/32

Yes

Yes

16k bits

16/32

Yes

Yes

32k bits

16/32

Yes

Yes

64k bits

16/32

Yes

Yes

128k bits

16/32

Yes

Yes

256k bits
(GLA-1016/1032)

16 only

No

No

256k bits
(GLA-1132)

32

Yes

Yes

512k bits
(GLA-1132 only)

32

Yes

Yes

1M bits
(GLA-1132 only)

32

Yes

Yes

2M bits
(GLA-1132 only)

16