beautypg.com

Trigger position, Trigger delay – GW Instek GLA-1000 Series User Manual User Manual

Page 56

background image

GLA User Manual

56

Trigger position

Toolbar icon

Sets the amount of pre-triggering data that GLA captures,
selectable from the pull-down menu.

Trigger

Pre-

triggering

Post-

triggering

0% means GLA starts capturing
data right after the triggering

condition (no pre-triggering data).

50% means half of the captured

data belongs to waveform prior to
the triggering condition (default).

100% means all of the captured
data belongs to waveform prior to

the triggering condition.

Trigger delay

Toolbar icon

Delays the data capturing range in reference to the Trigger
condition.

Trigger

Original range

Revised range

Delay

The amount of delay can be set
as Time…

Or number of clocks.

Trigger delay /
Trigger page

Trigger delay disables Trigger page function (page54), vice versa.