42

limitation

Using large amount of memory (256k bits for

GLA-1016/1032, 2M bits for GLA-1132)
imposes the following restrictions.

Only 16 channels, A0 ~ B7, available

Compression not available

Enable mode not available

Compression
mode

Toolbar Icon

When selected, GLA uses a lossless compression technique to
expand the signal record length. The compression ratio is maximum

255, automatically adjusted.

The maximum record length per channel is:

GLA-1016, 1032: 128kbit x 255 = 32Mbit
GLA-1132: 1024kbit x 255 = 256Mbit

The actual compression rate is displayed on the upper-right side of
the display (In the below case, 1.391).

Compression
limitation

Compression mode imposes the following
restrictions.

For 32 channel model, channel number is
reduced to 24, A0 ~ C7, regardless of the

memory size.

Enable mode becomes unavailable.

Enable mode

Opens the Enable function setup dialogue window. For details, see
the Trigger setup chapter, page57.

When
completed…

Press Apply/OK/Cancel button to save or cancel the setting.

Restore Defaults button setting: Internal clock, 100MHz/ 2k RAM