MULTI-PARAMETER TRANSMITTER USER’S MANUAL

Section 3 - Electrical Connections and Setup

Section 3 - Electrical Connections and Setup

Page 7

3.3 pH and ORP Differential Probe connections and setup

The drawing shows the connections for the Aquametrix Differential (5 wire) probe. The cable

should be run in a conduit separate from AC power wires, and via a separate conduit hole.

Figure 3.2 Connections for Differential (5 wire) pH or ORP probe

Once connected, step through the LCD menus to

select the probe in the order shown. The first two

steps may be skipped if the meter is already con-

figured for pH or ORP and a Differential Probe.

When using a pH probe, it is important to ensure

that the transmitter is reading the probe tempera-

ture correctly for accurate temperature compensa-

tion. The ORP probe does not require temperature

compensation, although the transmitter can dis-

play process temperature measured by the probe.

The factory temperature calibration is usually

accurate enough that no adjustments are neces-

sary.

Dwg# N104-42

CAUTION:

Always disconnect

the loop power

before wiring or

unwiring a sensor