Flow - utilities menu - manual test mode 7.2, Multi-parameter transmitter user’s manual – Aquametrix SHARK TX / TXP Transmitters User Manual User Manual
Page 68
Flow - Utilities Menu - Manual Test Mode 7.2
Page 66
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Flow - Utilities Menu - Manual Test Mode 7.2
DOWN
DOWN
DOWN
UP
DOWN
NOTE:
When the user exits the Manual Test Mode, the 4-20mA
output will remain in the Test Mode state until the operator
enters the run mode. The output will then revert back to
the previous On-line state.
The 4-20 mA output will track the
change in process value
RUN MODE
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
Press the key once
which will move the cursor
over the least digit of the
simulated process value.
Use the and keys to change
the simulated process value.
Manual Test Mode is used to simulate a process reading
in order to verify the correct response of the output.
When in the Manual Test Mode, the output is no longer
placed on hold as it is when in the rest of the menu.
Press the key once to
move the cursor to the RH side
of the display
Press the key to return to the
MANUAL TEST MODE menu
The 4-20 mA output will follow the process
value change.
DOWN
UP
DOWN
MANUAL TEST MODE
TOTAL 0
CALIBRATION
UTILITIES
TEST 65.00 GPM >
TEST 65.0 0 GPM >
14.4mA
14.4mA
TEST 65.0 0 GPM >
TEST 50.0 0 GPM >
TEST 50.00 GPM >
12.0mA
12.0mA
14.4mA
DOWN