MULTI-PARAMETER TRANSMITTER USER’S MANUAL

Section 3 - Electrical Connections and Setup

Section 3 - Electrical Connections and Setup

Page 9

3.5 Conductivity Cell (Contacting style) connections and setup

Figure 3.4 Connections for Conductivity Cells

Once connnected, step

through the LCD menus to

select the cell in the order

shown. The TEMP COMP

CURVE setup default is

1.8%/deg C. This is accept-

able for most process appli-

cations. If your process is

significantly different from

this, change the setting in

the TEMP COMP CURVE

menu.

The drawing shows the connections for the Aquametrix Conductivity Cells (Contacting style). The cable

should be run in a conduit seperate from the AC power wires, and via a seperate conduit hole. The cell

cable length should not exceed 300ft. (91 meters).

Dwg# N104-44

CAUTION:

Always disconnect

the loop power

before wiring or

unwiring a sensor