Ph - diagnostics menu - calibration data 4.13, Multi-parameter transmitter user’s manual – Aquametrix SHARK TX / TXP Transmitters User Manual User Manual
Page 28
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
pH - Diagnostics Menu - Calibration Data 4.13
pH - Diagnostics Menu - Calibration Data 4.13
Page 26
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
The Calibration Data menu is a series of read only screens which
allow the user to view the data collected during the last calibration.
RUN MODE
Press to view the first
Calibration Data screen.
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
1st POINT PH = 7.00
INPUT FROM PROBE = - 19 mV
Calculated efficiency based on 2
point calibration = 91 %
Calculated slope based on 2
point calibration = 59.16 mV/pH
Probe Temperature at which
calibration was performed = 19.4°C
2nd POINT PH = 4.00
INPUT FROM PROBE = + 143 mV
Calibration Mode
2 point calibration
Press to view the next
Calibration Data screen.
EFF 91%
>
59.16MV/PH
>
CALIBRATION DATA
CAL TEMP 19.4C >
CALIBRATION DATA
2P 4.00/ 143
>
CALIBRATION DATA
1P 7.00/ -19
>
MODE 2PT
>
CALIBRATION DATA
7.15pH 25.0C
DIAGNOSTICS
CALIBRATION
UTILITIES
SETUP
CAL EFFICIENCY
CAL SLOPE
CALIBRATION DATA
FIRMWARE REV.