MULTI-PARAMETER TRANSMITTER USER’S MANUAL

Conductivity - Utilities Menu - Manual Test Mode 6.4

Conductivity - Utilities Menu - Manual Test Mode 6.4

Page 50

DOWN

DOWN

DOWN

UP

DOWN

DOWN

UP

DOWN

Manual Test Mode is used to simulate a process reading
in order to verify the correct response of the output.
When in the Manual Test Mode, the output is no longer
placed on hold as it is when in the rest of the menu.

NOTE:

When the user exits the Manual Test Mode, the 4-20mA

output will remain in the Test Mode state until the operator

enters the run mode. The output will then revert back to the

previous On-line state.

The 4-20 mA output will track the
change in process value

RUN MODE

UP

NOTE

PRESS THE AND KEYS

TOGETHER TO GO IMMEDIATELY BACK TO

RUN MODE

DOWN

Press the key once to
move the cursor to the RH side
of the display

The 4-20 mA output will follow the process
value change.

Press the key to return to the
MANUAL TEST MODE menu

Use the and keys to change
the simulated process value.

Press the key once
which will move the cursor
over the least digit of the
simulated process value.

TEST

100 0 >

TEST

1000 >

1000uS 25.0C

CALIBRATION

UTILITIES

14.4mA

TEST 1310 >

TEST 131 0 >

14.4mA

MANUAL TEST MODE

12.0mA

14.4mA

TEST 131 0 >

12.0mA

DOWN