Yokogawa DA100 User Manual
Page 79

IM DA100-01E
4-10
Measurement range and accuracy
Note that RTD input is not possible for the DCV/TC/DI input module.
Reference operation conditions
temperature: 23
±2°C; humidity: 55 ±10% RH; supply voltage 90 to 250 VAC (AC power supply)
/10 to 32 VDC (DC power supply); supply frequency: 50/60
±1% (AC power supply); warm-up
time: at least 30 minutes; when operating, the system must not adversely affect the operation of
other equipment.
Input
Measuring(digital display)
DC Voltage
TC
(Note that Accuracy
of reference junction
compensation is not
considered).
RTD
High resolution RTD
Contact
20mV
60mV
200mV
2V
6V
20V
50V
R
S
B
K
E
J
T
L
U
N
W
KPvsAu7Fe
Pt100(1mA)
Pt100(2mA)
JPt100(1mA)
JPt100(2mA)
Pt50(2mA)
Ni100(1mA)
SAMA
Ni100(1mA)DIN
Ni120(1mA)
J263*B
Cu10 GE
Cu10 L&N
Cu10 WEED
Cu10 BAILEY
Pt100(1mA)
Pt100(2mA)
JPt100(1mA)
JPt100(2mA)
Voltage input
Contact input
-20.000 to 20.000mV
-60.00 to 60.00mV
-200.00 to 200.00mV
-2.0000 to 2.0000V
-6.000 to 6.000V
-20.000 to 20.000V
-50.00 to 50.00V
0.0 to 1760.0
°C
0.0 to 1760.0
°C
0.0 to 1820.0
°C
-200.0 to 1370.0
°C
-200.0 to 800.0
°C
-200.0 to 1100.0
°C
-200.0 to 400.0
°C
-200.0 to 900.0
°C
-200.0 to 400.0
°C
0.0 to 1300.0
°C
0.0 to 2315.0
°C
0.0 to 300.0K
-200.0 to 600.0
°C
-200.0 to 250.0
°C
-200.0 to 550.0
°C
-200.0 to 250.0
°C
-200.0 to 550.0
°C
-200.0 to 250.0
°C
-60.0 to 180.0
°C
-70.0 to 200.0
°C
0.0 to 300.0K
-200.0 to 300.0
°C
-140.00 to 150.00
°C
-70.00 to 70.00
°C
-140.00 to 150.00
°C
-70.00 to 70.00
°C
Off for a voltage of less
than 2.4V.
On for a voltage of 2.4V
or more.(TTL)
On/Off of contact
Type
Measurement accuracy
Maximum
resolution
±(0.05% of rdg + 5digits)
±(0.05% of rdg + 2digits)
±(0.05% of rdg + 2digits)
±(0.05% of rdg + 2digits)
±(0.05% of rdg + 2digits)
±(0.05% of rdg + 2digits)
±(0.05% of rdg + 2digits)
±(0.05% of rdg + 1°C)
However,R,S:0 to 100
°C,±3.7°C
100 to 300
°C,±1.5°C
B:400 to 600
°C,±2°C accuracy
less than 400
°C is not specified.
±(0.05% of rdg +0.7°C)
However,K attains an accuracy of
±(0.05% of rdg. +1°C) within the
range between -200 to -100
°C.
±(0.05% of rdg +0.5°C)
However,J and L attain an accuracy of
±(0.05% of rdg. +0.7°C) within the
range between -200 to -100
°C.
±(0.05% of rdg. +0.7°C)
±(0.05% of rdg. +1°C)
±(0.05% of rdg. +0.7K)
±(0.05% of rdg. +0.3°C)
±(0.05% of rdg. +0.3°C)
±(0.05% of rdg. +0.3°C)
±(0.05% of rdg. +0.3°C)
±(0.05% of rdg. +0.3K)
±(0.2% of rdg. +0.7°C)
±(0.05% of rdg. +0.3°C)
±(0.05% of rdg. +0.3°C)
±(0.05% of rdg. +0.3°C)
±(0.05% of rdg. +0.3°C)
Off for a voltage of less than 2.4V.
On for a voltage of 2.4V or more.
(TTL)
On/Off of contact
Measurement range
1
µV
10
µV
10
µV
100
µV
1mV
1mV
10mV
0.1
°C
0.1K
0.1
°C
0.1K
0.1
°C
0.01
°C
*
1
*
1
*
1
*
1
*
1
*
1
*
1
*
2
*
2
*
3
*
4
*
5
*
5
*
5
*
5
*
5
*
6
*
6
*
7
*
8
*
8
*
8
*
8
*
5
*
5
*
5
*
5
*1 : R,S,B,K,E,J,T:ANSI,IEC 584,DIN IEC 584,JIS C 1602-1981
*2 : L:Fe-CuNi,DIN43710,U:Cu-CuNi,DIN 43710
*3 : N:Nicrosil-Nisil,IEC 584,DIN IEC 584
*4 : W:W·5%RE-W·26%Re (Hoskins Mfg Co)
*5 : Pt50:JIS C 1604-1981,JIS C 1606-1986
Pt100:JIS C 1604-1989,JIS C 1606-1989,IEC 751,DIN IEC 751
JPt100:JIS C 1604-1981,JIS C 1606-1989
*6 : SAMA/DIN
*7 : McGRAW EDISON COMPANY
*8 : ranges to which accuracy applies:
Cu10 GE:-84.4 to 170.0
°C,
Cu10 L&N:-75.0 to 150.0
°C,
Cu10 WEED:-20.0 to 250.0
°C,
Cu10 BAILEY:-20.0 to 250.0
°C
Noise rejection
rejection by integration type A/D, lowpass filter, or moving averaging.
Cut-off frequency
50/60/10Hz
Input resistance
Min. 10M
Ω at 2V DC or lower, thermocouple range
Approx. 1M
Ω at 6V DC or higher.
4.2 Universal Input Module and DCV/TC/DI Input Module