Error message Page 65
Page 65

IM DA100-01E
3-6
Error message
No.
Error/Corrective Action
E3631
Communication not possible during logging.
Terminate Logger's communication before executing.
E3632
Communication not possible while monitoring data.
Terminate Logger's communication before executing.
E3633
Connected to unkown model.
Modify the system or change the settings.
E3636
Failed to send command.
Some items could not be set. Check the setup data.
E3637
Can't reconstruct.
Displayed when the DR stand-alone model is connected.
E3638
Initial balancing failed.
Check the input, initialize, and perform initial balancing again. If it fails again, servicing is needed.
E3639
Incorrect file name!
Check the file name and extension.
E3640
File operation failed.
Check the free space on the PC's hard drive. Check that the hard disk is operating properly.
E3641
These data were generated by another model.
Select a different file.
E3643
Failed to load data. Some data initialized.
Check the setup data.
E3644
Invalid channels are included in the copy source.
Check the copy source and copy destination settings.
E3645
Please enter an equation.
Enter an equation.
E3646
Unknown symbol is entered. Please change equation.
Check the setup data.
E3647
Invalid channel number is entered. Please change equation.
Check the channel number.
E3648
Incorrect use of parenthesis. Please change equation.
Check the setup data.
E3649
Grammatical error. Please change equation.
Check the setup data.
E3650
Character string is incorrect.
Check the setup data.
E3651
Invalid channel number.
Change the channel number.
E3652
For input range, please keep CH numbers "Left < Right".
Check the setup data.
E3653
Channel number is already being used!
Check the setup data.
3.2 Error Messages