Maintenance Model

7500ZA

Teledyne Analytical Instruments

82

7.4.1.3

R

EMOVING THE

M

EASURING

U

NIT

Refer to Fig. 7-3. For step 1 to 4, see 7.4.1.1, Pipe Cell Removal.

5. Disconnect and remove detector output cables from detector

output circuit board (No.12). Applying identification mark on
top of removed cable connector will ensure proper pin
assignment later.

6. Disconnect wiring to the 2-pin terminals of the IR light source

assembly and chopper motor pin connector from the PCB (No. 17).

7. Detach the six screws (No. 16) fastening the base plate (No. 3) to

remove the measuring unit. .

8. For assembly, reverse the disassembly procedures.

Note: Special care should be taken when assembling or

disassembling the measuring cell to avoid using excessive
force to the detector pipe or light source unit pipe. If the pipe
is deformed or damaged by excessive force, there is a
potential for gas leaks and this, in addition to creating a
potential toxic hazard, will result in inaccurate measurement.

7.4.2 Cleaning the Sample Cell

1. To clean the sample cell or IR transmission window, first clear

away any dirt using a soft brush and then lightly wipe using a
soft cloth. Do not use abrasive or paper towels.

Note: The window is fragile. Handle with care. Avoid scratching

the window. Do not rub the dirt from the window roughly.

2. If the window or the sample cell interior is very dirty, use a soft

line-free cloth moistened with absolute alcohol.

3. If the window is corroded, rub off any scale from the window

lightly with a soft cloth to which chrome oxide powder is
applied. If it is excessively corroded, it should be replaced.

4. When the cleaning operation is complete, assemble according to

the sample cell assembly/disassembly procedures. Assemble the
pipe carefully. If it becomes bent or damaged, replace it with a
new part.

5. Do not wash the sample cell components with water.