NDIR Gas Analyzer

Instalation

Teledyne

Analytical

Instruments

13

Without

O

2

Sensor

Built-in O

2

Sensor

External O

2

Sensor

measurement.

of 90 to 100% on
the measuring
range

90 to 100% on
the measuring
range.

90 to 100% on
the measuring
range.

Span gas for O

2

measurement

Gas with conc. of
90 to 100% on
the measuring
range or
atmospheric air
(21% O

2

).


(O

2

gas of 95 to

95.5 vol% for
reverse range O

2

measurement.)

O

2

gas at 1 to 2%

3.4.4 Internal Purging of the Instrument Housing

The internal housing does need not be purged except under the

following conditions:

1. A combustible gas component is contained in the sample gas.

2. Corrosive gas is contained in the atmospheric air at the

installation site.

3. The same gas as the sample gas component is contained in the

atmospheric air at the installation site.

In such cases, the analyzer housing should be purged with the

instrument air or dry N

2

.

The purge flow rate should be set to about 1L/min. The purge gas

must be free of dust and moisture.

3.4.5 Pressure at Sample Outlet

Pressure at the sample gas outlet should be atmospheric pressure.
Make sure there are no kinks or obstructions in the exhaust line.