NDIR Gas Analyzer

Specifications

Teledyne

Analytical

Instruments

99

Electromagnetic Compatibility ;

EN61010-1 : 2010 Safety requirements for electrical

equipment for measurement, control and
laboratory use.
“Installation Category II”
“Pollution Degree 2”

EN61326-1 : 2006

Electrical equipment for measurement,
control and laboratory use — EMC
requirements.

9.1.6 Requirements for Sample Gas

Flow rate:

0.5L / min ±0.2L / min

Temperature:

0 to 50˚C

Pressure:

10 kPa or less (Gas outlet side should be at

atmospheric pressure).

Dust:

100 µg/Nm3 or less in particle size of 0.3

µm or less.

Mist:

Not allowed

Moisture:

For sample gases NO, SO

2

, CO(0-200

ppm range): less than 2°C saturation point.
For most other sample gases: less than
standard room temperature saturation
point.

Corrosive component:

1 ppm or less

Standard gas for calibration:

1) Infrared-ray measurable component, standard O

2

Zero gas: Dry N

2

Span gas ; Each sample gas having

concentration 90 to 100% of its
measuring range recommended).

For external zirconia O

2

sensor use the

same calibration gas line: