Teledyne 7500ZA - Infrared Gas Analyzer User Manual

Page 106

background image

Specifications Model

7500ZA

Teledyne Analytical Instruments

90

O

2

External
Zirconia

0-5 vol.%

0-25 vol.%

Max. 5 components measurement

including O

2

.

For reverse range O

2

measurement, IR gas

measurement is not available (single range
O

2

only).

Measuring range ratio max. 1:10 (except

O

2

)

User adjustable ranges between the
specified minimum and maximum range.

Settable one range or two ranges, low

range (Range 1) and high range (Range 2).

See Table 1 for possible combinations of

components and ranges.

Display Indication:

Digital indication in 4 digits (LCD panel

with LED back light)

Instantaneous value of each component

Instantaneous value after O

2

correction

(only in NO, SO

2

, CO measurement with

O

2

)

Average value after O

2

correction (only in

NO, SO

2

, CO measurement with O

2

)

O

2

average value

Analog Output Signals:

4 to 20mA DC or 0 to 1V DC, isolated

internally from circuit and ground. Output
lines are non-isolated; 12 outputs max.

Allowable load 550Ω for 4 to 20mA DC