NDIR Gas Analyzer

Settings and Calibration

Teledyne

Analytical

Instruments

73

6.8 Calibration

6.8.1 Zero Calibration

This function is used for the zero

point adjustment. A proper zero gas,
suitable for the application, should be
used. Refer to section 3.4 “Sampling”.

1. From the Measurement screen,

press the ZERO key to display the
Manual Zero Calibration screen.

2. Select the Ch (component) to be

calibrated using the UP or DOWN
key and then press the ENT key,
and zero gas will be admitted to
the analyzer.

Note: For the Ch (components)

where “both” is set in the

calibration range setting

(

see

section 6.2.3), zero calibration
is carried out on both ranges.

3. Wait until the indication stabilizes

on the zero gas then press ENT.
The zero calibration will be carried
out on the highlighted range(s).

Note: On Ch (components) for which

“AR” is set (see Section 6.1.1
Setting Range Switch Mode),
the cursor automatically moves
to the next range selected in
“Setting of Auto Calibration
Component/Range” (6.2.4),
and calibration is carried out
on that range.

To close the Zero Calibration screen

or to abort the entry midway, press the
ESC key. The previous screen will return.