NDIR Gas Analyzer

Settings and Calibration

Teledyne

Analytical

Instruments

69

6.7.3 Calibration Log screen

Calibration history can be displayed from this screen. Information

available includes sensor input value, concentration value, and the date
when zero/span calibration was performed for the last 10 calibration
events. They are logged according to component.

To clear an event, move the

cursor to Clear Calibration Log and
press the ENT key, and the
calibration log is cleared completely.

Other items appearing on the

Calibration Log screen are:

Z1 : Zero calibration (Z) of Range 1

S1 : Span calibration (S) of Range 1

Cnt : Value of measuring detector at

the time of calibration

Con : Concentration value displayed

before calibration

YDMH: Time of calibration event