NDIR Gas Analyzer

Specifications

Teledyne

Analytical

Instruments

91

Allowable load 100KΩ for 0 to 1V DC

 Refer to Table2 for the channel No. of

displayed values and analog output
signals.

Analog Input Signal:

For signal input from externally installed

O

2

sensor.

Signal

requirement:

(1) Signal from Zirconia O

2

sensor

(TYPE: ZFK7)

(2) 0 to 1V DC from an O

2

sensor. Input

section is not isolated. This feature is
present when an O

2

sensor is not built

in.

* Externally installed O

2

sensor must be

purchased separately.

Digital Output: (option)

1C contact (24V DC/1A, resistive load)

max.15 outputs

Instrument error, calibration error,
range identification, auto calibration
status, solenoid valve drive for auto
calibration, High/Low limit alarm
contact output.

All relay contacts are isolated mutually

and from the internal circuit.

Digital Input: (option)

Voltage contact (supply 12-24VDC

(15mA Max.)) Max. 9 inputs

Remote range change over, auto
calibration remote start, remote hold,
average value reset, Isolated from the
internal circuit with photocoupler.