Settings and Calibration

Model 7500ZA

Teledyne Analytical Instruments

60

6.6 Parameter Setting

From the Parameter Setting screen you can set or adjust various

parameters of the instrument such as time, key lock, etc. Settable items
are as follows:

Description of setting items

• Current Time : Current year, month, date, day of the week, hour,

and minute setting. (The display appears in this order.)

• Note: The clock backup time is 2 days. If power is turned on

after being off for 2 days or longer, the time must be set again.

• Key Lock : Invalidates any key operation except canceling the

key lock.

• Output Hold : Sets whether Calibration Output is held or not, and

the holding value setting.

• Response time : Sets the response time of electrical system.

• Average Period : Sets the moving average time.

• Backlight Timer : Sets automatic OFF of the backlight of display

unit and its timer.

• Contrast : Adjusts contrast of the LCD.

• Maintenance Mode : A password protected entry screen used to

access the instrument’s maintenance mode.

* For the maintenance mode, see Section 6.7.

1. Enter the “Parameter

Setting” screen from the user
mode, and the display shown
at right appears. Use the UP
or DOWN key to move the
cursor to the desired item
and then press ENT.