NDIR Gas Analyzer

Settings and Calibration

Teledyne

Analytical

Instruments

51

Auto calibration status contact output is closed during auto

calibration (NO side), and is open in other cases.

Example:

Start Time

Sun

12:00

Cycle 1

day

Flow Time

Zero

350 sec

CH 1 Span

350 sec

CH 2 Span

350 sec

CH 3 Span

350 sec

CH 4 Span

300 sec

CH 5 Span

300 sec

Ex. Time

300 sec

ON/OFF ON

Thus, with the above settings:

Setting Range:

Cycle:

1-99 hours or 1-40 days (default value—7days)

Flow Time:

60-90 sec (default value—300 sec)