background image

NDIR Gas Analyzer

Settings and Calibration

Teledyne

Analytical

Instruments

59

Auto Zero Calibration Screen

Example

With Ch1 and Ch2 set to enable for auto

calibration (see Section 6.2.4):

• Zero calibration

A message, “Zero cal.” blinks at Ch1
and Ch2.

Caution: During auto zero calibration,

the only key operations allowed are key lock
ON/OFF and “Auto Zero Calibration Stop.”

When the key lock is set at ON, even the “Auto
Zero Calibration Stop” cannot be used.

To forcibly stop “auto zero calibration”, set the
key lock to OFF and then execute “Auto Zero
Calibration Stop.”