Maintenance Model

7500ZA

Teledyne Analytical Instruments

78

7.4 Cleaning the Sample Cell

Entry of dust or water in the sampling cell contaminates the interior

of the cell resulting in drift. If dirty or contaminated, it must be cleaned.
Check the sampling device, especially the filter, to prevent the cell from
being dust or mist contamination.

Caution: Maintenance

should be performed only by

properly trained and qualified personnel.

In addition to the following steps, the user must
adhere to any local and regional regulations.

7.4.1 Sample Cell Disassembly

There are two kinds of sampling cells used in the Model 7500ZA:,

block cells (cell length: 4 mm, 8 mm, 16 mm, 32 mm) and pipe cells
(cell length: 64 mm, 125 mm, 200 mm and 250 mm).

A two-component analyzer may incorporate both sampling cells in

one optical unit. In this case, detach the pipe cell and then block cell (see
Fig. 7-3).