Installation Model

7500ZA

Teledyne Analytical Instruments

18

• Used when the external zirconia O

2

sensor or the external O

2

sensor is ordered. It is not required or used if the analyzer has a
built-in O

2

sensor.

• Connect the dedicated connector (accessory) to the output of the

external zirconia sensor or the external O

2

sensor (shipped

separately).

• When using an external O

2

sensor, the input signal should be 0 to

1 VDC with respect to the O

2

full scale of the analyzer. The O

2

concentration display, output, and O

2

correction can then be

performed correctly.

Note: The O

2

sensor input is not isolated. TAI recommends that it

be isolated when an external O

2

analyzer is installed

remote from the analyzer. The zirconia O

2

sensor should

be installed at a location that is as close to the analyzer as
possible.

The figure below shows how to connect the O

2

signal to the

dedicated connector (accessory).