NDIR Gas Analyzer

Settings and Calibration

Teledyne

Analytical

Instruments

47

6.3.3 Peak Alarm Setting

When the peak number of times CO concentration exceeds the upper

limit value during measurement reaches the set number, an alarm is provided.

The peak alarm and this setting screen

appear only when an option is added.

1. Press the MODE key while in the

Measurement mode, and the User
mode appears.

2. Using the UP or DOWN keys move

the cursor to “Setting of Peak
Alarm”. Then press ENT.

3. In the “Peak Alarm Setting”

selection screen that appears, point
the cursor to any item you want to set
using the UP or DOWN key. Then
press ENT.

4. Enter numeric values using the UP or

DOWN keys. The setting will be
saved after pressing the ENT key.

Description of Setting Items

Peak Alarm : ON/OFF of peak alarm

Alarm Value : If measuring value exceeds

the set alarm value, a peak
counter increments by 1
time.

Alarm Count : When the alarm value is

exceeded this many times per
hour, the peak count alarm is
activated (closed).

Hysteresis :

To prevent possible
chattering when the
measuring value may exceed
the set peak concentration by
only 1 time, the peak count
has an allowance in the
hysteresis width.