• Fit of all Standards - The standard curve will be determined by using each individual
standard replicate when calculating the standard curve. For example, if you run duplicates
of a seven-point standard curve, the software will calculate the standard curve by using 14
points. Applies only to quantitative analysis.
• Mean of Replicates - The standard curve will be determined by averaging the individual
standard replicates when calculating the standard curve. For example, if your un duplicates
of a seven-point standard curve, the software calculates the standard curve by using 14
points. Applies only to quantitative analysis.
Cancel- Returns you to the main Batches tab.
Next - Click to advance to the Analytes tab.
Log Tab
This tab displays a log of the activity that occurred during the acquisition of the selected
batch.
FIGURE 37.
Log Tab
This tab displays the following information about each activity: