Luminex xPONENT for MAGPIX User Manual
Page 33

FIGURE 7.
Batch Options Tab
Batch Options. This section displays the following options:
• Allow running a batch with expired reagents - Allows batches to run with expired
reagents.
• Allow running a batch if XY temperature is not in range - Allows batches to run if the
XY temperature is not in range.
• Allow running a batch if the instrument is not calibrated or verified - Allows batches to
run if the instrument is not calibrated.
• Prompt for reacquiring of the last errored wells for an aborted partial batch - Allows
users to choose to reacquire aborted wells when resuming a stopped/partial batch or to
start with the next unacquired well.
Batch Settings This section displays the following options:
• Calibration expiration days - Type the desired number of days a system calibration
remains active before expiring. The default number of days is seven.
• Analysis display decimal places - Type the number of digits the system displays. The
default number of digits is three.
xPONENT for MAGPIX