Classification and Reporter Fluorochromes
MagPix beads in the calibration kit are used to autofocus the camera and calibrate the CL1,
CL2, and RP1 channels. The beads in the verification kit are a mix of 6 different regions that
cover the range of the 50-plex map. Both calibration and verification beads are triple-dyed,
and the fluorescence signal of these dyes enables classification of each bead set.
TABLE 1.
MAGPIX Active Bead Regions (by Region)
Region
Region
Region
MC10012
MC10013
MC10014
MC10015
MC10018
MC10019
MC10020
MC10021
MC10022
MC10025
MC10026
MC10027
MC10028
MC10029
MC10030
MC10033
MC10034
MC10035
MC10036
MC10037
MC10038
MC10039
MC10042
MC10043
MC10044
MC10045
MC10046
MC10047
MC10048
MC10051
MC10052
MC10053
MC10054
MC10055
MC10056
MC10057
MC10061
MC10062
MC10063
MC10064
MC10065
MC10066
MC10067
MC10072
MC10073
MC10074
MC10075
MC10076
MC10077
MC10078
Fluidics 1 and Fluidics 2
Although it undergoes a wash step in between wells, the probe can be susceptible to carry-
over from well-to-well. Fluidics 1 contains one bead set. Fluidics 2 contains a buffer solution
and a different control bead. The function of this maintenance procedure is to measure how
much (as a percentage) of the first bead set in Fluidics 1 is found in the well where Fluidics 2
has been loaded.
Sample Volume
Sample volumes range in size from 20 µL to 200 µL. Ensure that approximately 25 µL more
than the sample volume remains in the well after aspiration. This amount may vary
depending on the type of plate used. Your sample volume must be large enough to prevent
aspirating air into the fluid line when acquiring sample, and small enough to prevent spill-over