Managing sample lists – Luminex xPONENT for MAGPIX User Manual
Page 72

• Dilution - Contains the following dilution options:
• 1:2 - Halves the standard from each previous iteration.
• 1:10 (Log) - Computes a value of one-tenth of the standard from each previous iteration.
• 1/2 Log - Creates a 1:3.16 dilution, or half of each 1:10 (Log) from each previous
iteration.
Alternatively, type a number for your own dilution factor.
• Apply Dilution - Applies the dilution selected in the Dilution list.
NOTE: Click a column header to re-sort the display.
NOTE: Click the Reagent column header to re-sort the order from the
highest number standard to standard number one. This is useful for
applying dilutions in which the last standard is the highest standard.
Assay Control Information - Lists the selected control reagents. The list displays the
Reagent, Name, Lot Number, Expiration, and Manufacturer. Existing control lot
information can be applied or new information can be typed manually.
• Apply Ctrl Lot - Opens the Select Lot dialog box. Select a lot from the list and then click
OK.
• Show Concentration - Expected, Low, and High set the expected, lowest, or highest
acceptable concentration of the analyte in the sample.
• Apply Values - Applies a value down or across the list of analytes.
Cancel - Returns to the Batches tab.
Back - Returns to the previous tab.
Next - opens the Plate Layout tab.
Managing Sample Lists
Use the Samples page to create new samples, view the details of a sample list, or to create
a new batch. Samples can also be added with a Lab Information System (LIS). To learn more
about how to use a LIS with xPONENT, see
on page 11.
Using xPONENT
59