beautypg.com

Revive after storage routine – Luminex xPONENT for MAGPIX User Manual

Page 102

background image

FIGURE 41.

Sample Probe Height Adjustment

1. Well D6

2. Well D1

3. Well RB1

Revive After Storage Routine

After you have adjusted the sample probe height, run the Revive After Storage (Luminex)
routine.

1. Open the Maintenance page, then the Cmds & Routines tab.

2. Select Revive After Storage (Luminex) from the drop-down list. The Revive After

Storage routine performs the following commands:

Prime

Rinse

Alcohol Flush

Rinse

After the Revive After Storage routine is complete, run the Calibration/Verification routine.

Performing System Maintenance

89