Revive after storage routine – Luminex xPONENT for MAGPIX User Manual
Page 102
FIGURE 41.
Sample Probe Height Adjustment
1. Well D6
2. Well D1
3. Well RB1
Revive After Storage Routine
After you have adjusted the sample probe height, run the Revive After Storage (Luminex)
routine.
1. Open the Maintenance page, then the Cmds & Routines tab.
2. Select Revive After Storage (Luminex) from the drop-down list. The Revive After
Storage routine performs the following commands:
• Prime
• Rinse
• Alcohol Flush
• Rinse
After the Revive After Storage routine is complete, run the Calibration/Verification routine.
Performing System Maintenance
89