beautypg.com

Equivalent point determination (ip methods), Titra lab – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 96

background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 96

Equivalent point
determination
(IP methods)

An equivalent point of a titrated species/titrant reaction is determined
in two ways depending on the titrant addition mode (continuous or
incremental) selected:

Continuous addition of the titrant (Continuous IP method):

acquisition of the titration curve E (electrode) = f (volume of titrant
dispensed),

calculation of the first derivative curve with smoothing process
(method parameter: smoothing criterion),

determination of the second derivative curve,

determination of the inflection points on the first derivative curve
from the minimum and maximum of the second derivative curve,

determination of the tangents to the first derivative curve for these
inflection points,

determination of the equivalent point as the intersection of the two
tangents.

Figure 17: Determination of an equivalent point - Continuous IP method

An equivalent point is determined and plotted live on the titration
curve. The method can be stopped after a given number (1 to 4) of
equivalent points has been detected.

E or pH