Titra lab – Hach-Lange TITRALAB 840_845 Reference Manual User Manual
Page 206

Titra
Lab
®
TitraLab 840 and 845 Reference Manual
Page 206
The following tables give the conversion factor to use in the calculation equation.
If the result is in volume, the result given will be the volume Vi added to reach the equivalence
point. This volume will be converted in ml or µl if asked for.
R = V *c
with:
Sample volume Vs in ml:
Result
c (if ml conversion)
c (if µl conversion)
in ml by default
1
1000
“C
smp
”
“C
reag
”
“c” if “Vs” in “ml”
eq/l = N
eq/l = N
1
eq/l = N
meq/l
0.001
eq/l = N
mol/l = M
x
eq/l = N
mmol/l
0.001 x
meq/l
Same formula as C
smp
= eq/l but multiplied by 1000
mol/l = M
Same formula as C
smp
= eq/l but multiplied by 1 / y
mmol/l
Same formula as C
smp
= eq/l but multiplied by 1000 / y
g/l = mg/ml
Same formula as C
smp
= eq/l but multiplied by M
smp
/ y
mg/l
Same formula as C
smp
= eq/l but multiplied by M
smp
* 1000 / y
mol
Same formula as C
smp
= eq/l but multiplied by Vs * 0.001 / y
mmol
Same formula as C
smp
= eq/l but multiplied by Vs / y
g
Same formula as C
smp
= eq/l but multiplied by M
smp
* Vs * 0.001 / y
mg
Same formula as C
smp
= eq/l but multiplied by M
smp
* Vs / y
µg
Same formula as C
smp
= eq/l but multiplied by M
smp
* Vs * 1000 / y