TitraLab 840 and 845 Reference Manual
Page 163
Running a
method
(steps 6 to 9)
see "Running a method (preliminary steps)", page 162
.
6.
User ID and sample temperature.
7.
Sample ID and Test portion quantity.
8.
Rinse electrodes then dip electrodes in test 1 and confirm.
9.
Measurements start after a Start timer delay (if programmed).
If
Sample Run Start = Shortcut
is selected in
the Setup > Configuration menu, screens of steps 7 to 9 are
not proposed. So, it is important to dip the electrodes into
the sample before running the method at step 6 !
Refer to "Sample Run Start/Run End (Standard/Shortcut)",
page 173
Refer to "Running a method (steps 10 to 12)", page 164
.
Entry of the User ID (if User ID
= Yes has been selected),
see "User ID (Yes/No)", page 194
Entry of the sample temperature (if
Temperature = Entered
has been selected),
see "Temperature Probe/ Fixed at
Entry of the Sample ID and Test
portion quantity (titration methods
only).
Refer to "Start timer", page 184
Refer to "Run window", page 161
.