Create electrode, Create method, Titra lab – Hach-Lange TITRALAB 840_845 Reference Manual User Manual
Page 66

Titra
Lab
®
TitraLab 840 and 845 Reference Manual
Page 66
Create electrode
1.
Enter the Electrode window.
2.
Press 4 then 1.
3.
Select the electrode function,
.
4.
Press
✓ in the ID field.
5.
Select From = Other.
The option From = Catalogue allows you to create an
electrode from a list of Radiometer Analytical electrodes.
6.
Enter the electrode name (up to 16 alphanumeric characters).
7.
In the Confirm creation screen:
•
For pH, mV or ISE electrodes only; select the electrode type.
Refer to "Electrode type", page 89
.
•
For combined pH or single pH electrodes; enter the internal pH
(pH int) of the electrode.
•
For combined pH, Metal/Redox or ISE electrodes or for a
Conductivity electrode; select if the electrode has a built-in
temperature sensor or not.
•
For reference electrodes or combined pH, Metal/Redox, ISE
electrodes; enter the potential of the reference (in mV) versus the
Standard Hydrogen Electrode.
Refer to "Potential versus SHE", page 131
8.
Press 1 to create the electrode. The Edit electrode menu is
displayed. Enter the electrode definition parameters.
Create method
1.
Enter the Main window.
2.
Press 5 then 1.
3.
Press
✓ in the New method screen.
4.
Enter a method name (up to 17 alphanumerical characters).
5.
Press 1 to create the method. The Edit method menu is
displayed.
6.
Enter the method parameters.