Fill burette, Flush burette, Format (printouts) – Hach-Lange TITRALAB 840_845 Reference Manual User Manual
Page 101: Titra lab

Titra
Lab
®
TitraLab 840 and 845 Reference Manual
Page 101
Fill burette
1.
Enter the Burette functions menu:
2.
Select the burette in the Address field,
.
3.
Press 1 Fill. The burette is filled at maximum speed using the
installed reagent. Approximate filling time: 25 s.
Flush burette
A flush rinses the burette, tubings and delivery tip thoroughly. It is
recommended to flush the burette on a daily basis.
The Titration Manager carries out a flush after each installation or rea-
gent replacement procedure.
To run a flush:
1.
Enter the Burette functions menu.
2.
Select the burette in the Address field (TitraLab 845 only).
3.
Press 3 Flush.
The flush procedure runs as follows:
•
Empty,
•
Fill 4 ml (*) then empty (5 times).
(*) 1 ml for a 10 ml burette and 10 ml for a 50 ml burette.
The length of the “Flush” procedure depends on the volume of the
burette.
You can also flush simultaneously all installed burettes of a
TitraLab 845,
see "Global flush burettes", page 103
.
Format
(printouts)
Format = Listing
The whole report is printed in one operation.
Format = Page by page
The printer waits until a preset number of lines have been collected
then prints one page (this number is set by the Nb line per
page parameter),
see "Nb lines per page (printouts)", page 124
The printing format applies for automatic printouts (at the end
of a test) or manual printouts (by pressing key Print).
Access:
Setup Menu > Configuration
Refer to "Printouts", page 136
.