3. Create new method or Edit a pre-programmed one
Create the measurement or titration method to be used for the analyses. Enter the parameters
required to calculate the results,
see "Programming methods", page 25
.
When you have finished programming, select the method or pre-programmed application,
If you are using a sample changer, edit the sequence and program the sample stack,
.
4. Check icons
The following icons indicate the exact state of your working system:
Refer to "Electrode icons", page 87
Refer to "Reagent icons", page 149
Sunny icons are needed in order to run the selected method.
If Cloudy/Stormy/Question mark icons are displayed in the Reagent/Electrode window
press 1 to activate the “Check” command. The Titration Manager will automatically
guide you through the operations required to solve the errors encountered.
B.Running methods
To run a method or sequence,
see "Working in Routine mode", page 18
Sunny icon:
Everything is OK. Run the method or sequence.
Cloudy icon:
Electrode/reagent calibration required within 12 or 24 hours.
Stormy icon:
Electrode/reagent calibration date elapsed or reagent(s) not installed.
Question mark:
Programming error.