Remove burette, Remove method from a sequence, See "remove – Hach-Lange TITRALAB 840_845 Reference Manual User Manual
Page 153: Titra lab

Titra
Lab
®
TitraLab 840 and 845 Reference Manual
Page 153
Remove burette
If no reagent is installed
1.
Enter the Burette functions menu.
2.
Press 5 Remove burette then press
✓. The piston is moved to its
lowest position (filling operation).
3.
Check the burette data, remove the burette and press
✓.
If a reagent is already installed
The reagent is removed and the burette flushed before removal.
1.
Enter the Burette functions menu:
2.
Press 5 Remove burette then press
✓. The piston is moved to its
uppest position (emptying operation).
3.
Remove the reagent bottle and press
✓. The burette is filled then
emptied.
4.
Install a rinse bottle containing demineralised water and press
✓.
.
5.
Remove the rinse bottle and press
✓. The burette is filled,
emptied then filled again. At the end, the piston stops at its lowest
position.
6.
Check the burette data, remove the burette and press
✓.
Remove method
from a
sequence
1.
Press 3 Sequence/Sample stack in the Main window.
If this option is not available, enter the Setup menu > Configura-
tion and declare a SAC80 or SAC850 Sample Changer (Supervi-
sor access only).
2.
Press 3 Edit sequence.
3.
Select the method to be removed using the LEFT/RIGHT arrow
keys.
4.
Press 3 Delete.
5.
Press 1 Remove method then
✓ to confirm or press Esc to leave
the screen without removing the method.
The method is removed from the sequence but is not deleted.
To remove all the methods from a sequence,
.