Working in Routine mode
A.Access methods
A Routine operator has access to all screens for checking purposes
B.Running methods
When working in “ROUTINE” mode, it is necessary to install your titration system according to
the selected method or sequence, prior to running a method or sequence.
1. Select the method or sequence
Refer to "Select method", page 175
Refer to "Select sequence", page 175
2. Connect the electrodes
Refer to "Electrode connection", page 86
.
3. Check icons
Refer to "Check icons", page 17
.
Depending on the icons displayed, the Titration Manager will automatically guide you through
the steps necessary to run the analysis, see below:
a. Install reagents(s)
Check that the burette is installed, if not,
see "Install burette", page 107
.
see "Install reagent", page 108
.
b. Calibrate electrode(s)
Now, run the calibration,
see "Calibrate electrodes", page 55
c. Calibrate reagent(s) or Enter titre
Now, run the calibration or enter the titre.
Refer to "Reagent calibration", page 145
Refer to "Enter titre", page 92
d. Run the method or the sequence
Refer to "Running a method (preliminary steps)", page 162
Refer to "Running a sequence with a sample changer", page 166
.