Reagent addition volume, Reagent calibration, Refer to "reagent – Hach-Lange TITRALAB 840_845 Reference Manual User Manual
Page 145: Titra lab

Titra
Lab
®
TitraLab 840 and 845 Reference Manual
Page 145
Reagent
addition volume
Option available with TitraLab 845.
Volume of reagent to be added.
Enter in:
Edit method > Parameters (if Addition = Yes)
Edit reagent > Calibration parameters (if Addition = Yes)
Range available:
0.01 ml to 999 ml
Reagent
calibration
1.
Select the method which uses the reagent to be calibrated.
2.
If a Question mark "?" is present in the Reagent and/or Electrode
tabs, it means that the method needs to be programmed - a rea-
gent or an electrode is missing. Review programming in Supervi-
sor mode.
Refer to "Programming method", page 138
3.
Install the reagent(s),
see "Install reagent", page 108
.
4.
Connect the electrode,
see "Electrode connection", page 86
.
5.
Press 1 Calibrate/Enter titre in the Reagent window.
6.
Select the titrant from the list.
7.
Press 1 Run calibration and follow the messages on the display.
8.
Measurements start.
The titre ±standard deviation is calculated at the end of the
calibration cycle. The calibration results are saved with the
reagent.
To consult the calibration results,
Standard ID.
Time elapsed since titration started
Temperature measured or entered
Potential measured
Titrant ID
Test number
Volume of titrant dispensed in ml
Titrant burette addition speed
Potential (mV) = f (volume dispensed (ml))